The best way to prevent leakage is to ensure that the retention balloon is adequately inflated and properly positioned in the rectum at all times during the use of Flexi-Seal™ FMS.
Review carefully the "Insertion of device" section of the Directions for Use and make sure that all the instructions for proper insertion are followed. After ensuring that the balloon is properly positioned, take note of the position indicator line relative to the patient’s anus. Observe changes in the location of the position indicator line to determine movement of the balloon in the rectum.
If the position indicator line moves up, ie., closer to the anus, this may indicate that there is a need for gently pulling the catheter to bring the balloon where it should be.
If the position indicator line moves down, ie., away from the anus, this may indicate that the device may expel. In such cases, the balloon should be deflated, removed, and reinserted as described in the "Insertion of Device" section of the Directions for Use.